TB 9-6625-2102-24
(6) Set TI FREQ/MULT(Hz) to switch settings and frequency dial position for each
row in table 19. If measured period using frequency counter does not indicate within limits,
perform b (2) below.
NOTE
The remaining tests are done using period measurements.
Some checks are lengthy.
Test instrument
Frequency counter indications (ms)
Test instrument
FREQ MULT (Hz)
switch
Frequency dial
Adjustments
settings
positions
Min
Max
10
2.0
48.07
52.08
R51 (R)
10
1.0
94.33
106.38
1
1.0
943.3
1063.8
1
2.0
480.7
520.8
0.1
2.0
4807.0
5208.0
0.01
2.0
48000.0
52000.0
b. Adjustments
(1) Perform adjustments listed in table 18 and repeat as required for in-tolerance condition.
(2) Perform adjustments listed in table 19 and repeat as required for in-tolerance condition.
16. Main Generator Balance and Zero
a. Performance Check
(1) Connect TI FUNCTION OUT (50 :) to audio analyzer INPUT HIGH using a
50 : feedthrough termination.
(2) Initiate audio analyzer and set to measure level with DC low-pass filter on.
(3) Set main FUNCTION switch to DC and adjust AMPLITUDE control fully ccw.
(4) If audio analyzer is not within limits in table 20, perform b (l) below.
Audio analyzer indications
(V)
Min
Max
-.020
.020
(5) Disconnect TI from audio analyzer.
(6) Connect TI FUNCTION OUT (50 :) to oscilloscope Vertical 1 input.
(7) Connect MODULATION GENERATOR OUT (600 :) to EXT MOD IN.