TB 9-6625-2102-24
indications
(%)
Min
Max
49.5
50.5
(7) If oscilloscope does not indicate within duty cycle limits in table 13, perform
b (2) below.
indications
(%)
Min
Max
45
55
b. Adjustments
(1) Adjust R36 (fig. 2) until oscilloscope indicates a duty cycle between 49.5% and 50.5% (R).
NOTE
R36 is effective on positive portion of square wave.
(2) Adjust R23 (fig. 2) until oscilloscope indicates a duty cycle between 45% and 55% (R).
NOTE
R23 is effective on negative portion of square wave.
(3) Repeat this performance check and adjustment procedure as necessary for a best
in-tolerance condition.
10