TB 9-6625-2117-35
14.
NOTE
Do not perform power supply checks if all other parameters are
within tolerance.
a. Performance Check.
Connect digital multimeter to
-15 V
test point as
shown in
figure 6 for SN
499999 and
below or
figure 7 +15V for SN
500000 and
above. If
digital
voltmeter does not indicate between -14.888 and -15.112 V (+14.888 and +15.112 V),
perform b below.
Figure 6. Adjustment and test point location -
top view - SN
499999 and below.
b. Adjustments . Adjust 15 VOLT R1122
(fig. 6) for SN 499999 and below
(+15 VOLT R1940 (fig. 7) for SN 500000 and above) for -15 V (+15 V) indication on digital