TB 9-6625-2124-35
(a)
SHIFT key, DATA ENTRY 002.
(b)
GHz key, DATA ENTRY 80.
(c)
M2 key, DATA ENTRY 00.
↑ key, DATA ENTRY 00.
(d)
↑ key, DATA ENTRY 00.
(e)
↑ key, DATA ENTRY 00.
(f)
(8) The FREQUENCY display will indicate 2C83 and the FREQUENCY/TIME
display will indicate 00; if not, repeat (3) through (7) above.
(9) Adjust A8R22 (fig. 6) for multimeter indication of -7.000 V dc (R).
(10) Press ↓ key two times. The FREQUENCY display will indicate 2C81.
(11) Press function keys and enter corresponding data on keyboard as listed in (a)
through (d) below:
↓ key, DATA ENTRY 0 BKSP.
(a)
↑ key, DATA ENTRY 0 BKSP.
(b)
↑ key, DATA ENTRY 0 BKSP.
(c)
↑ key, DATA ENTRY 0 BKSP.
(d)
24