• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
HF and LF-3 dB Point

Calibration Procedure For Differencial Amplifiers Am-6881/U (Type 4A13) And Am-6786/U (Type 7A22) Page Navigation
  10    11    12    13    14  15
img
.
TB 9-6625-2132-24
Table 10. High and Low Frequency -3dB Point
Test instrument
Function/arbitrary
HF And LF
generator frequency limits
-3db POINT
Min
Max
switch settings
.3 MHz
270 kHz
330 kHz
.1 MHz
90 kHz
110 kHz
30 kHz
27 kHz
33 kHz
10 kHz
9 kHz
11 kHz
3 kHz
2.7 kHz
3.3 kHz
.3 kHz
270 Hz
330 Hz
.1 kHz1
90 Hz
110 Hz
10 Hz2
8.8 Hz
11.2 Hz
.1 kHz
88 Hz
112 Hz
1 kHz
880 Hz
1120 Hz
10 kHz
8.8 kHz
11.2 kHz
1Return
HF-3dB POINT switch is left at 1 MHz position
2Starts
LF-3dB POINT switch positions.
16