Figure 5. Dropout waveforms.

Calibration Procedure For Data Error Analyzer Model 1645A Page Navigation
  6    7    8    9    10  11
img
. .
TB 9-6625-2138-24
a. Performance Check
(1) Connect multimeter between +5 V test point (fig. 6) and chassis ground.
Multimeter indication will be between 4.75 and 5.25 V dc.
Figure 6. Test instrument - bottom interior view.
(2) Connect multimeter between +15 V test point (fig. 6) and chassis ground.
Multimeter indication will be between 14.5 and 15.75 V dc.
NOTE
Top cover of TI must be removed to gain access to LOGIC
PROBE CONNECTOR (fig. 1).
(3) Connect  multimeter  between  center  conductor  of  LOGIC  PROBE
CONNECTOR and chassis ground. Multimeter indication will be between 4.75 and 5.25 V.
b. Adjustments. No adjustments can be made.
14. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.