TB 9-6625-2155-35
Table 2. Minimum Specifications of Equipment Required - Continued
Minimum use
Manufacturer and model
Common name
specifications
(part number)
MEASURING RECEIVER
Range: 2.0 to 18 GHz
Hewlett-Packard, Model
Range: +8 to -74.5 dBm
8902A w/sensor, Hewlett
Accuracy: +.5 dB
Packard, Model 11792A, and
Range: AM 0 to 80%
microwave converter, model
Accuracy: 2% at 1 kHz
11793A
Range: FM .05 to 100 kHz
Accuracy: 2% at 1 kHz
Deviation: 12 kHz
Range: 0 to 40 V dc
Fluke, Model 8840A/AF 05/09
Accuracy: 0.01%
(AN/GSM-64D)
Range: 5.0 V at 100 ns
(OS-303/G)
Accuracy: 3.0%
Range: Period 10 s
LeCroy Model 9210 with
plug-in, LeCroy Model 9211
SYNTHESIZED SIGNAL
Range: Output level +8 dBm
Anritsu, Model 68369NV
(AN/USM-677)
SPECTRUM ANALYZER
Range: 2.0 to 18 GHz at +10 to 60 to dBm
of the span/div
Range: Span 500 Hz to 20 MHz
Accuracy: 5%
TIME/FREQUENCY
Frequency: 1 MHz
Datum, Model ET6000-75
WORKSTATION
Accuracy: 5 parts in 10
-10 per day
(13589305)
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual and TM 11-6625-3143-
40 for this TI.
d.
When indications specified in
paragraphs 8 through 14 are not within
tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 14. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.