TB 9-6625-2155-35
Table 6. Harmonic Distortion Check - Continued
Spectrum analyzer
Test
indication
instrument
(dB)
Harmonic
frequency
Harmonic
frequency
(GHz)
maximum (≤)
14
4.6667
.33
-35
14
9.3333
.66
-35
16
5.3333
.33
-35
16
10.6667
.66
-35
18
6
6 GHz Sub
-35
18
12
12 GHz Sub
-35
resolution bandwidth to 10 kHz, and span to 1 MHz.
(10) Using the spectrum analyzer, verify that all non-harmonic spurious signals
are ≤ 50 dBc TI frequencies listed in table 7.
Table 7. Non-harmonic Spurious Signal Level
Test instrument frequency
(GHz)
Spurious signal level
2.500 000
< -50 dBc
3.000 000
< -50 dBc
3.500 000
< -50 dBc
4.500 000
< -50 dBc
5.000 000
< -50 dBc
5.500 000
< -50 dBc
6.500 000
< -50 dBc
(11) Reduce all outputs to minimum and disconnect equipment.
b. Adjustments. No adjustments can be made.
a. Performance Check