Table 8. Residual THD + N

Calibration Procedure For Distortion Analyzer, Ts-4084/G Page Navigation
  8    9    10    11    12  13
img
. . . . . . .
TB 9-6625-2170-24
(3) Adjust R1030 (fig. 1) for minimum TI indication (R).
(4) Adjust R1023 (fig. 1) for minimum TI indication (R).
(5) Repeat (3) and (4) above, as necessary, due to interaction.
(6) Position signal generator FREQUENCY Hz controls for 800 Hz.
(7) Adjust R5025 (accessible through hole in bottom cover) for minimum TI
indication (R). If R5025 has no effect, adjust to center position.
15. Power Supply
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
a. Performance Check.
(1) Connect multimeter INPUT HI to TP 3041 +5V and LO to TP 3044 GND (fig. 2).
(2) If multimeter does not indicate between 4.98 and 5.51 V dc, adjust R3047 +5V
ADJ (fig. 2) until multimeter indicates 5.25 r0.1 V dc (R).
16. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.