SECTION III. CALIBRATION PROCESS FOR TS-402/U AND TS-402A/U
SECTION IV. CALIBRATION PROCESS FOR CN-1000/G, HEWLETT-PACKARD, MODEL 350D, AND NDL, MODEL CO-710

Calibration Procedure For Attnuators, Ts-402/U, Ts-402A/U, Cn-1000/G, Model 350D, Ndl, Model Co-710, And General Radio, Types 1450-Tar And 1450-Tb Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . .
TB 9-6625-2181-24
Figure 1. Frequency response.
8. Attenuator Accuracy
a. Performance Check
(1) Energize equipment and allow sufficient time for warm-up.
(2) Adjust calibrator controls for 10.00 V, 1 kHz output.
(3) Note and record indication on true rms voltmeter.
(4) Set TI 600  DB LOSS switch to 1. Adjust calibrator amplitude until true rms
voltmeter indicates value recorded in (3) above. Calibrator will indicate between 8.7096
and 9.1201 V ac.
(5) Repeat technique of (4) above for switch settings listed in table 3.
Table 3. Attenuator Accuracy 1 kHz
Test instrument
Calibrator output indications
600  DB LOSS
(V ac)
switch settings
Min
Max
2
7.7625
8.1283
3
6.9183
7.2444
5
5.4954
5.7544
10
3.0903
3.2359
20
0.9772
1.0233
40
0.0977
0.1023
5