SECTION IV. CALIBRATION PROCESS FOR CN-1000/G, HEWLETT-PACKARD, MODEL 350D, AND NDL, MODEL CO-710
Frequency Response

Calibration Procedure For Attnuators, Ts-402/U, Ts-402A/U, Cn-1000/G, Model 350D, Ndl, Model Co-710, And General Radio, Types 1450-Tar And 1450-Tb Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. . . . . .
TB 9-6625-2181-24
11. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Set all TI DB switches to 0.
b. Connect equipment as shown in figure 2.
Figure 2. Attenuator range and accuracy.
12. Attenuator Range and Accuracy
a. Performance Check
.
(1) Set decade resistor to 600
(2) Energize equipment and adjust calibrator for an indication of 45 V dc on
multimeter.
(3) Set TI 0 to 10 DB switch to 1. Multimeter will indicate between 39.53325 and
40.68764 V dc.
(4) Repeat technique of (3) above for switch settings listed in table 5.
CHANGE 1
7