SECTION II. EQUIPMENT REQUIREMENTS
Equipment Setup

Calibration Procedure For Radio Frequency Power Test Set, An/Usm-491 (Model 4200 Rf Microwattmeter And Series 4200-6E Sensor) Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . .
TB 9-6625-2185-24
Table 2. Minimum Specifications of Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
Fluke, Model 5720A (5720A)
CALIBRATOR
Range:  0.09 to 10 V dc
(p/o MIS-35947)
Accuracy:  r 0.125%
FREQUENCY COUNTER
Range:  100 kHz to
Fluke, Model PM6681/656
50 MHz
(PM6681/656)
Accuracy:  r0.25%
MULTIMETER
Range:  0 to 5.2 V dc
Agilent, Model 3458A (3458A)
Accuracy:  r0.002%
Agilent, Model E12-432A
POWER METER
Range: -10 to +7 dBm
(MIS-30525) w/thermistor mount,
Accuracy:  r(0.95% at
Agilent, Model 478A-H75
50MHz)
(7915907) or 8478B (8478B)
1
Biddle-Gray, Model 71-631
RESISTANCE STANDARD
Range:  554 k:
2
Accuracy:
(7910328)
SYNTHESIZER/LEVEL
Range:  -34 to +10 dBm
Agilent, Model 3335AOPT 001-
GENERATOR
Accuracy:  r0.05 dBm
K06 (MIS-35938)
SYNTHESIZED SIGNAL
Range: 0.05 to 18 GHz
Anritsu, Model 68369NV
GENERATOR
Accuracy:  r0.5%
(68369NV)
Range  r10 dBm
1
Two required.
2
Combined accuracy of calibrator and resistance standard: 0.125%.
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by
common name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual and TM 11-6625-
3164-14 for this TI.
d. When indications specified in paragraphs 9 through 10 are not within tolerance,
perform the power supply check prior to making adjustments.  After adjustments are
made, repeat paragraphs 9 through 10. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
4