TB 9-6625-2185-24
(13) If calculated PO is not between 0.993 and 1.007 mW, perform b below.
b. Adjustments
(1) Remove two screws (A) and remove front panel strip (fig. 1).
Figure 1. Adjustment locations - bottom view.
(2) Adjust AR4 (fig. 1) so that calculated PWR REF output power is
1 mW as calculated in (7) through (12) above. (R)
CHANGE 1 7