TB 9-6625-2192-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Cavity Type Frequency Meters (300 MHz to 40 GHz) (General). The
manufacturers' manuals were used as the prime data sources in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. Variations among models are described in the text.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the microwave technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Manufacturer (part number)
Performance specifications
Harris, PRD, Type 587A
Range: 300 MHz to 1 GHz
Accuracy: -0.5% 300 to 400 MHz
0.2% 400 MHz to 1 GHz
Range: 0.96 to 4.2 GHz
FEL, Model WDA-940
Accuracy: 0.22% 0.96 to 1 GHz
Hewlett-Packard, Model 536A
0.17% 1 to 4.2 GHz
Harris PRD, Type LS-518
Wavetronics, Model 3301A-90
FXR, Model N410
Range: 1 to 4 GHz
Accuracy: 0.1%
FEL, Model WDA-3712
Range: 3.7 to 12.4 GHz
Harris PRD, Type CX-518
Accuracy: 0.17%
Hewlett-Packard, Model 537A
Wavetronics, Model 3303A-00
Hewlett-Packard, Model G532A
Range: 3.95 to 5.85 GHz
Waveline, Model 398DR
Accuracy: 0.065%
General Microwave, Model N608A
Range: 3.95 to 8.20 GHz
Accuracy: 0.1%
FXR, Model N414A
Range: 3.95 to 11.0 GHz
Accuracy: 0.1%
Hewlett-Packard, Model J532A
Range: 5..30 to 8.20 GHz
Accuracy: 0.065%