TB 9-6625-2192-24
Table 8. Frequency Dial Accuracy (General Microwave, Model N608A and FXR, Model N414A) - Continued
Test instrument frequency dial settings
Microwave frequency counter indications
(GHz)
(GHz)
General microwave model
N608A
FXR, model N414A
Min
Max
---
8.5
8.49150
8.50850
---
9.0
8.99100
9.00900
---
9.5
9.49050
9.50950
---
10.0
9.99000
10.01000
---
10.5
10.48950
10.51050
---
11.0
10.98900
11.01100
Table 9. Frequency Dial Accuracy (Hewlett-Packard, Model J532A)
Test instrument frequency
Microwave frequency counter indications (GHz)
dial settings (GHz)
Min
Max
5.3
5.296555
5.303445
5.5
5.496425
5.503575
6.0
5.996100
6.003900
6.5
6.495775
6.504225
7.0
6.99545
7.004550
7.5
7.495125
7.504875
8.0
7.994800
8.005200
8.2
8.194670
8.205330
Table 10. Frequency Dial Accuracy (FR1 26/U; FXR, Models X41OB and X41OBS; Hewlett-Packard,
Model X532B; Waveline, Model 698DR; and Wavetronics, Model 3007-02)
Test instrument frequency
Microwave frequency counter indications (GHz)
dial settings (GHz)
Min
Max
8.2
8.19344
8.20656
8.5
8.49320
8.50680
9.0
8.99280
9.00720
9.5
9.49240
9.50760
10.0
9.99200
10.00800
10.5
10.49160
10.50840
11.0
10.99120
11.00880
11.5
11.49080
11.50920
12.0
11.99040
12.00960
12.4
12.39008
12.40992
Table 11. Frequency Dial Accuracy (Hewlett-Packard, Model P532A)
Test instrument frequency
Microwave frequency counter indications (GHz)
dial settings (GHz)
Min
Max
12.4
12.3876
12.4124
12.5
12.4875
12.5125
13.0
12.9870
13.0130
13.5
13.4865
13.5135
14.0
13.9860
14.0140
14.5
14.4855
14.5145
15.0
14.9850
15.0150
15.5
15.4845
15.5155
7