Frequency Accuracy (300 MHz to 18 GHz) - Continued
Figure 3. Frequency accuracy (1 8 to 40 GHz) - equipment setup.

Calibration Procedure For Cavity Type Frequency Meters (300 Mhz To 40 Ghz) (Gneral) Page Navigation
  6    7    8    9    10  11  12    13  
img
. . . . . . . .
TB 9-6625-2192-24
Figure 2. Wavetronics, Model 3301 A - parts illustration.
NOTE
Power meter indication must remain at dip indication for (17)
below.
(17) Adjust tuning shaft c (fig. 2) for frequency dial indication of 4.00 GHz indication.
NOTE
Frequency of cavity must not change for (17) above.
(18) Tighten two setscrews e (fig. 2) on drive gear f (fig. 2).
(19) Replace TI cover on base assembly.
(20) Replace and tighten 4 cap screws d (fig. 2).
(21) Position knob b (fig. 2) on tuning shaft c (fig. 2) and tighten three setscrews a
(fig. 2).