Figure 3. Frequency accuracy (1 8 to 40 GHz) - equipment setup.

Calibration Procedure For Cavity Type Frequency Meters (300 Mhz To 40 Ghz) (Gneral) Page Navigation
  8    9    10    11    12  13
TB 9-6625-2192-24
(6) Repeat (3) through (5) above for remaining frequency settings listed in
selected table.
b. Adjustments. No adjustments can be made; however, a correction chart may be
prepared listing actual microwave frequency counter indication at each dial setting of
selected table.
10. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
14