Figure 2. Wavetronics, Model 3301 A - parts illustration.
Final Procedure

Calibration Procedure For Cavity Type Frequency Meters (300 Mhz To 40 Ghz) (Gneral) Page Navigation
  7    8    9    10    11  12  13  
img
. . . .
TB 9-6625-2192-24
9. Frequency Accuracy (18 to 40 GHz)
a. Performance Check
(1) Connect equipment as shown in figure 3.
Figure 3. Frequency accuracy (1 8 to 40 GHz) - equipment setup.
(2) Set TI frequency dial to first frequency setting in table selected in 7 a above.
(3) Adjust synthesized signal generator output controls for a convenient indication
on power meter.
(4) Adjust synthesized signal generator frequency controls for a maximum dip
((decrease) in power meter indication for some TIs, power meter indication will peak
(increase)).
(5) Microwave frequency counter indications will be within limits specified in
selected table.