TB 9-6625-2192-24
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
POWER SPLITTER NO. 1
Frequency range: 298.5 MHz to 18 GHz Weinschel, Model 1870A
Output tracking between ports:
(7916839)
298.5 MHz to 2 GHz: 0.15 dB
2 to 8 GHz: 0.2 dB
8 to 18 GHz: 0.25 dB
POWER SPLITTER NO. 2
Wiltron, Model K241() (K241())
1
GHz)
Output tracking between ports:
18 to 40.048 GHz (40 GHz)1: 0.6 dB
SYNTHESIZED SIGNAL
Frequency range: 298.5 MHz to 40.048
Anritsu, Model 68369NV
1
GHz(40 GHz)
(68369NV)
Output range: At least +3 dBm
Frequency accuracy: Accuracy of
microwave frequency counter
1
Standards limitations.
5. Accessories Required. The accessories listed in table 3 are issued as indicated in
paragraph 4 above, and are used in this calibration procedure. When necessary, these
items may be substituted by equivalent items, unless specifically prohibited.
Common name
Description (part number)
Adapter waveguide to coaxial
(10519423) (8 to 12.4 GHz) 2 each
(7923163) (12.4 to 18 GHz) 2 each
(13335442-2) (18 to 26.5 GHz) 1 each
(13335442-4) (26.5 to 40 GHz) 1 each
Cable assembly flexible RF
(18876-1) (12 inch) or
(18876-2) (24 inch)
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.