TB 9-6625-2197-40
Figure 8. Transmitter Power - Equipment Setup.
NOTE
pushbutton and repeat (4) above several times to ensure
maximum pulse amplitude display.
(5) Connect equipment as shown in figure 8, CONNECTION B.
(6) Press TI LAMP TEST pushbutton.
(7) Measure peak power from SUM connector.
Peak power meter will indicate
between + 2 and + 6 dBm.
NOTE
Set cursor delay on peak power meter to coincide with the top
of one of the pulses. Figure 9 is included as an example of a
graphical display from the peak power meter. Pressing graph
twice will initiate AUTOSCALING capability of the peak power
meter and provide a menu which can then be changed by
selecting menu options with the up and down arrow.
NOTE
Pressing TI LAMP TEST pushbutton several times will
provide viewing of multiple pulses on the peak power graphical
display.
17