TB 9-6625-2197-50
(4) Press and release LAMP TEST pushbutton. Amplitude difference between
pulses displayed on oscilloscope will not be greater than 10 percent of average amplitude.
NOTE
MODE 4 SIMULATED WORD (from the SUM connector)
displayed on oscilloscope consist of 37 pulses with number 5
missing.
NOTE
Press oscilloscope STORAGE ERASE pushbutton and repeat
(4) above several times to ensure maximum pulse amplitude
display.
(5) Connect equipment as shown in figure 8, CONNECTION B.
(6) Press TI LAMP TEST pushbutton.
(7) Measure peak power from SUM connector.
Peak power meter will indicate
between + 2 and + 6 dBm.
NOTE
Set cursor delay on peak power meter to coincide with the top
of one of the pulses. Figure 9 is included as an example of a
graphical display from the peak power meter. Pressing graph
twice will initiate AUTOSCALING capability of the peak power
meter and provide a menu which can then be changed by
selecting menu options with the up and down arrow.
Figure 9. Example of
peak power meter display.
NOTE
Pressing TI LAMP TEST pushbutton several times will
provide viewing of multiple pulses on the peak power graphical
display.
20