TB 9-6625-2197-50
(2) Set control box PWR switch to ON and set RPT and SEQ switches to OFF.
(3) Position frequency counter controls to measure pulsed RF.
(4) Press TI LAMP TEST pushbutton. Frequency counter will indicate between
1029.794 and 1030.206 MHz.
NOTE
Due to changing RF levels, repeat (4) above several times for
best results.
b. Adjustments. No adjustments can be made.
NOTE
Insertion loss of circulator and cable between TI and radar test
set at 1090 MHz must be known before the following procedure
is performed.
a. Performance Check
Figure 11. Receiver sensitivity -
equipment setup.