Table 1. Calibration Description - Continued
Table 2. Minimum Specifications of Equipment Required

Calibration Procedure For Radar Test Set An/Tpm-25A (Ts-3394A) Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . .
TB 9-6625-2211-24
Table 1. Calibration Description - Continued
Test instrument parameter
Performance specifications
Mode 4 challenge and GTC trigger
Range:
Amplitude:
5V
Width: 0.5 to 0.8 Ps
Delay:  182.5 and 386.5 Ps
Accuracy:
Amplitude:
r0.5 V
Width: r0.05 Ps (0.5 Ps)
r0.02 Ps (0.8 Ps)
Delay:  r0.06 Ps
RF output frequency
Range:
CW:
1090 MHz
SWP
r5 MHz: 1085 to 1095 MHz
SWP
r15 MHz:
1075 to 1105 MHz
Accuracy:
CW:
r0.109 MHz
SWP
r5 MHz: r0.1 MHz
SWP
r15 MHz:
r0.3 MHz
Output power
Range: 0 to -90 dBm
Accuracy:
r 1 dBm
Modulator and demodulator
Range:
Width: 0.45 Ps r0.05 Ps
Risetime:
.075 Ps
Falltime:
0.125 Ps
Accuracy:
Width: 0.05 Ps
Risetime:
r.025 Ps
Falltime:
r0.075 Ps
Input power
Range: -12 dBm to +3 dBW and +18 to +33 dBW
r 1 dBW1
Accuracy:
Input accuracy:
Range:
SWP
r5 MHz: 1025 to 1035 MHz
SWP
r15 MHz:
1015 to 1045 MHz
Accuracy:
SWP
r5 MHz: r0.2 MHz
SWP
r15 MHz:
r0.6 MHz
Challenge video
Range:
Amplitude:
5 and 20 V
Width: 0.8 Ps
Spacing: 3.0 to 21.0 Ps
Accuracy:
Amplitude:
r0.5 V (5.0 V)
r4.0 V (20.0 V)
Width: r0.05 Ps
Spacing: r0.1 Ps
1Procedure
limitation (limited to r 1 dB of standards used).
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
4