TB 9-6625-2211-24
NOTE
Verify the proper CAL FACTORS are loaded for the measuring
receiver's sensor module.
a. Remove TI protective cover as required for adjustments.
c. Connect autotransformer to a 115 V ac source and adjust for 115 V output.
d. Position controls as listed in (1) through (23) below:
MEASUREMENT PRF RANGE switch to X1K.
(1)
MEASUREMENT FUNCTION SEL switch to PRF REPLY TRIG.
(2)
MEASUREMENT 75:/2k: switch to 75:.
(3)
MEASUREMENT DEMOD VID LEVEL control to midrange.
(4)
MEASUREMENT FREQ MEAS control fully cw.
(5)
REPLIES MODULATION SEL switch to SIF.
(6)
REPLIES REPLY WIDTH SELECT switch to 0.45.
(7)
REPLIES SIF REPLY CODE switches to 7700.
(8)
REPLIES SUB PULSE SEL switch to OFF.
(9)
(10) REPLIES SUB PULSE POS switch to 0.
(11) REPLIES RANGE DELAY SEL (USEC) switches to 0001.
(12) REPLIES GATING PASS switches to 00.
(13) REPLIES GATING INHIB switch to 00.
(14) REPLIES M4 JAMMING switch to 0.
(15) PRT SEL (USEC) (500 USEC MIN) switches to 0500.
(16) SIF CHAL VID SIF MODE SEL switch to OFF.
(17) SIF CHAL VID 5V/20V switch to 5V.
(18) TRIG SEL DCD MODE SEL switch to OFF.
(19) TRIG SEL INT/DCD/EXT switch to INT.
(20) SIG GEN FUNCTION switch to FIXED FREQ.