TB 9-6625-2215-24
a. Performance Check
(1) Press 80 MHz RANGE and 10 Hz RESOLUTION pushbuttons and release
CHK pushbutton.
(2) Connect function generator output to TI 80 MHz jack, using termination.
until measuring system indicates a stable count of applied frequency. Function generator
output will not exceed limit specified in first row of table 9.
(4) Repeat technique of (3) above at function generator frequencies listed in table 9.
Output amplitude of function generator at each frequency will not exceed limit specified in
Function generator output
Frequency
Amplitude limits
(Hz)
(mV)
≤ 50
50
≤ 25
100
≤ 25
1
1000
≤ 25
10000
≤ 25
100000
≤ 25
1000000
≤ 25
10000000
≤ 25
25000000
≤ 25
50000000
≤ 50
80000000
Press TI 100 Hz RESOLUTION button.
1
(5) Remove signal input from 80 MHz jack and connect to 525 MHz jack without
termination; press 525 MHz RANGE pushbutton.
until measuring system indicates a stable count of applied frequency. Function generator
output will not exceed limit specified in first row of table 10.
(7) Repeat technique of (6) above at frequencies listed in table 10.
If function
generator output exceeds limits specified in table 10, perform b below.