Figure 2. Test Instrument - (HP 5304B) - Bottom View.
Table 13. DC Voltage.

Calibration Procedure For Measuring System, Td-1209/U (Model 5300A); 50 Mhz Universal Counter, Td-1211/U (Model 5302A); 10 Mhz Counter (Model 5301A); 500 Mhz Counter (Model 5303A); 525 Mhz Counter (Model 5303B); Timer/Counter (Model 5304A) And Multimeter/Counter (Model 5306A) Page Navigation
  17    18    19    20    21  22  23    24    25  
img
. . . . . .
TB 9-6625-2215-24
SECTION VIII
CALIBRATION PROCESS FOR MULTIMETER/COUNTER (HP 5306A)
28. Preliminary Instructions
a. The instructions outlined in paragraphs 28 and 29 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in tables 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified all controls and controls settings refer to the TI.
29. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Remove bottom cover from TI.
NOTE
Do not handle yellow printed circuit board except at extreme
edges.  Fingerprints on this board may degrade instrument
performance.
b. Mount a calibrated TD-1209/U (Hewlett-Packard, Model 5300A) measuring system
on TI and orient the interconnected units upside down.
c. Connect measuring system to autotransformer and connect autotransformer to a
115 V ac source.
d. Adjust autotransformer output for 115 V.
e. Adjust measuring system SAMPLE RATE control cw to on position and allow
1 hour for warm-up.
21