Table 13. DC Voltage.
Resistance

Calibration Procedure For Measuring System, Td-1209/U (Model 5300A); 50 Mhz Universal Counter, Td-1211/U (Model 5302A); 10 Mhz Counter (Model 5301A); 500 Mhz Counter (Model 5303A); 525 Mhz Counter (Model 5303B); Timer/Counter (Model 5304A) And Multimeter/Counter (Model 5306A) Page Navigation
  19    20    21    22    23  24  25  
img
. . . . . . . .
TB 9-6625-2215-24
Figure 3. Test instrument - (HP 5306A) - Bottom View.
31. Ac Voltage
a. Performance Check
(1) Connect lead between TI HI and LO terminals.
(2) Press FUNCTION ACV pushbutton and RANGE pushbuttons as listed in
table 14. If measuring system indications are not as specified in table, perform b (1) below.
(3) Remove lead from HI and LO terminals.
Table 14. ACV Zero.
Test instrument
Measuring system
RANGE pushbutton settings
indications
(V)
10
0.0000 (0.0020)
100
00.000 (00.050)
1000
000.00 (000.50)
(4) Connect calibrator to TI HI and LO terminals.
(5) Press RANGE 10 V pushbutton.
(6) Set calibrator output to 9.000 V at 40 Hz. If TI does not indicate within limits
specified in first row of table 15, perform b (2) below.
(7) Repeat technique of (5) and (6) above, using values listed in table 15. If TI does
not indicate within limits specified in table 15, perform b (2) below.
23