SECTION VIII. CALIBRATION PROCESS FOR MULTIMETER/COUNTER (HP 5306A)
Figure 3. Test instrument - (HP 5306A) - Bottom View.

Calibration Procedure For Measuring System, Td-1209/U (Model 5300A); 50 Mhz Universal Counter, Td-1211/U (Model 5302A); 10 Mhz Counter (Model 5301A); 500 Mhz Counter (Model 5303A); 525 Mhz Counter (Model 5303B); Timer/Counter (Model 5304A) And Multimeter/Counter (Model 5306A) Page Navigation
  18    19    20    21    22  23  24    25  
img
. . . . . . . . .
TB 9-6625-2215-24
30. Dc Voltage
a. Performance Check
(1) Connect leads between TI GUARD and LO terminals and between LO and
HI terminals.
(2) Press FUNCTION DCV and RANGE 10 V pushbuttons and release FAST
pushbutton. Measuring system will indicate 0.0000 (0.0003).
(3) Press RANGE 100 V pushbutton. Measuring system will indicate 00.000 (00.003).
(4) Press RANGE 1000 V pushbutton. Measuring system will indicate 000.00 (000.30).
(5) Press RANGE 10 V pushbutton and remove jumper from HI and LO terminals.
CAUTION
Do not exceed maximum voltages indicated on front panel of TI.
(6) Connect calibrator to TI HI and LO terminals.
(7) Set calibrator output to +9.0000 V dc.  If TI does not indicate within limits
specified in first row of table 13, perform corresponding adjustment listed in table 13.
Table 13. DC Voltage.
Calibrator
Test instrument
Limits
Output
RANGE
Adjustment
(Vdc)
Min
Max
10 V
b (1)
9.0000
8.9970
9.0030
10 V
b (2)
-9.0000
-9.0030
-8.9970
100 V
b (3)
90.000
89.970
90.030
1000 V
b (4)
900.00
898.83
901.17
(8) Repeat technique of (7) above using calibrator outputs and setting TI to range
settings listed in table 13.  If TI does not indicate within limits specified in table 13,
perform corresponding adjustment listed in table.
b. Adjustments
(1) Adjust +V ADJ R24 (rear panel) to obtain +9.0000 V dc indication on
measuring system.
(2) Adjust -V ADJ R1 (rear panel) to obtain -9.0000 V dc indication on
measuring system.
(3) Adjust A2 R18 (fig. 3) to obtain indication of +90.000 V dc on measuring system.
(4) Adjust A2 R24 (fig. 3) to obtain indication of +900.000 V dc on measuring system.