Calibration Procedure For Pulse Generator Type S-52 Page Navigation
  4    5    6    7    8  9
TB 9-6625-2218-24
18. Pulse Amplitude and Risetime
a. Performance Check
(1) Connect PRETRIG OUT (85ns) to AUX TRIG IN of oscilloscope.
(2) Connect PULSE OUTPUT to CH 1 of oscilloscope.
(3) Waveform on oscilloscope will have an amplitude of at least 200 mV.
(4) Measure risetime using standard measurement technique.
Risetime as
indicated on oscilloscope will be 350 ps or less.
b. Adjustments. No adjustments can be made.
19. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA Label/Form in accordance with TB 750-25.