Frequency Response
Table 7. Input Attenuators

Calibration Procedure For Spectrum Analyzer Model 8558B Page Navigation
  20    21    22    23    24  25  26    27    28    29    30  
img
. . .
TB 9-6625-2235-24
(2) Adjust TUNING control for an indication of 500 MHz on FREQUENCY MHz
readout and press FREQUENCY CAL pushbutton.
(3) Connect equipment as shown in figure 10, connection A.
Figure 10. Frequency response - equipment setup.
(4) Adjust signal generator frequency controls to 500 MHz and RF output controls
for a reference level signal peak on 6th division above graticule baseline.
(5) Record power meter indication (reference).
(6) Slowly adjust signal generator frequency and TI TUNING control from 80 to
1500 MHz while maintaining power meter reference level established in (5) above. The
signal level will not exceed l division of 6th horizontal graticule line.
(7) Connect equipment as shown in figure 10, connection B.
(8) Set RESOLUTION switch to 100 kHz and FREQ SPAN/DIV switch to 1 MHz.
(9) Adjust TUNING control for 80 MHz indication on FREQUENCY MHz readout
and press FREQUENCY CAL pushbutton.
(10) Adjust synthesizer/level generator frequency controls to 80 MHz and adjust
amplitude controls to reference established in (4) above on TI.
26