SECTION I. IDENTIFICATION AND DESCRIPTION
SECTION II. EQUIPMENT REQUIREMENTS
SECTION III. CALIBRATION PROCESS
Equipment Setup
Equipment Setup - Continued
Figure 1. Major assembly locations.
Calibrator Output Accuracy
Figure 2. Calibrator levels - equipment setup.
Figure 3. Adjustment location.
Figure 4. Sweep time - equipment setup.
Table 3. Sweep Time
Figure 5. A8 sweep generator - component and test point locations (SN prefix 2215A and below
Figure 7. Span width and frequency accuracy - equipment setup.
Figure 8. Frequency span accuracy measurement for 9th spectral line.
Span Width and Frequency Accuracy
Table 5. Frequency Accuracy - Continued
Figure 9. A7 frequency control - adjustment locations.
Span Width and Frequency Accuracy - Continued
Resolution Bandwidth Accuracy
Resolution Bandwidth Accuracy - Continued
Resolution Bandwidth Accuracy - Continued
Resolution Bandwidth Accuracy - Continued
Table 6. Bandwidth Adjustments
Frequency Response
Figure 10. Frequency response - equipment setup.
Table 7. Input Attenuators
Figure 11. Reference level - equipment setup.
Reference Level Accuracy
Figure 12. Log amplifier adjustments - equipment setup.
Reference Level Accuracy - Continued
Figure 13. A14 log amplifier assembly adjustment locations for SN prefix 2436A.
Reference Level Accuracy - Continued
Reference Level Accuracy - Continued
Figure 14. A14 log amplifier - adjustment locations (SN prefix 2142A and below).
Reference Level Accuracy - Continued
Residual FM Test
Figure 15. Residual FM.
Final Procedure