Table 3. Sweep Time
Figure 7. Span width and frequency accuracy - equipment setup.

Calibration Procedure For Spectrum Analyzer Model 8558B Page Navigation
  7    8    9    10    11  12  13    14    15    16    17  
img
. . . . .
TB 9-6625-2235-24
Figure 5. A8 sweep generator
Figure 6. A8 sweep generator
- component and test point
assembly -
locations (SN prefix 2215A
component locations
and below
(SN prefix 2332A)
10. Span Width and Frequency Accuracy
a. Performance Check
(1) Connect equipment as shown in figure 7.
(2) Position controls as listed in (a) through (g) below.
(a) FREQ SPAN/DIV switch to 20 MHz.
(b) RESOLUTION BW (RES BW) switch pushed in to OPTIMUM (coupled).
(c) INPUT ATTEN switch to 0 dB (OPTIMUM INPUT switch to -40 dBm).
(d) REFERENCE LEVEL switch to -20 dBm.
(e) 10 dB/DIV pushbutton pressed in.
(f) TIME/DIV switch to AUTO.
13