Figure 2. Calibrator levels - equipment setup.
Figure 4. Sweep time - equipment setup.

Calibration Procedure For Spectrum Analyzer Model 8558B Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
. . . . .
TB 9-6625-2235-24
(6) Adjust A9L4 (fig. 3) for a maximum signal amplitude indication on TI display (R).
Figure 3. Adjustment location.
9. Sweep Time
a. Performance Check
(1) Connect equipment as shown in figure 4.
10