Vertical Deflection
Table 4. Channel 2 Vertical Deflection Accuracy

Calibration Procedure For Oscilloscope Os261A(V)1U (Type 475 Option 7), Os261B(V)1U (Type 475 Option 4), Os261C(V)1U (Type 475, Option 4/7), And Type 4751 (Type 4F75A) Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . .
TB 9-6625-2240-24
Figure 1. Left side view.
(10) Perform steps a through c below:
(a) Use technique of (b) and (c) below for TI VOLTS/DIV switch settings and
oscilloscope calibrator VOLTAGE outputs listed in table 3.
(b) Rotate CH 1 POSITION knob to center trace on TI crt.
(c) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton
to obtain 6 divisions of vertical display. Oscilloscope calibrator err display will indicate
within limits specified in table 3; if not, perform b below.
Table 3. Channel 1 Vertical Deflection Accuracy
Test instrument
Oscilloscope calibrator
Oscilloscope calibrator
VOLTS/DIV
Err display limits
Voltage output
(%)
switch settings
2
m
12
mV
3
51
m
30
mV
3
10
m
60
mV
3
20
m
120
mV
3
50
m
.3
V
3
.1
V
.6
V
3
.2
V
1.2
V
3
.5
V
3.0
V
3
1
V
6.0
V
3
2
V
12.0
V
3
5
V
30.0
V
3
V1
10
60.0
V
3
1For
Tektronix, Type 475A.
6