Table 12. Delay Time and Differential Measurement
Table 14. High Frequency Compensation Adjustments

Calibration Procedure For Oscilloscope Os261A(V)1U (Type 475 Option 7), Os261B(V)1U (Type 475 Option 4), Os261C(V)1U (Type 475, Option 4/7), And Type 4751 (Type 4F75A) Page Navigation
  8    9    10    11    12  13  14    15    16    17    18  
img
. . . . . . . . . . .
TB 9-6625-2240-24
NOTE
Interaction occurs between R936 and R938 (fig. 3). Readjust as
necessary for best in-tolerance condition.
(3) Adjust R936 (fig. 3) until start of sweep is aligned with the leading edge of the
displayed time marker (R).
(4) Set DELAY TIME POSITION dial to 1.00.
(5) Adjust C1023 (fig. 2) to align time marker with center vertical graticule line (R).
(6) Set DELAY TIME POSITION dial to 9.00.
(7) Readjust C1023 (fig. 2) for best in-tolerance condition between DELAY TIME
POSITION dial readings of 1.00 and 9.00 to align time marker with center vertical
graticule line (R).
(8) Set DELAY TIME POSITION dial to 1.00.
(9) Adjust C1025 (fig. 2) to align time marker with center vertical graticule line (R).
(10) Set DELAY TIME POSITION dial to 9.00.
(11) Readjust C1025 (fig. 2) for best in-tolerance condition between DELAY TIME
POSITION dial readings of 1.00 and 9.00 to align time marker with center vertical
graticule line (R).
11. Risetime and Bandwidth
a. Performance Check
(1) Press HORIZONTAL DISPLAY A (A LOCKS KNOBS for SN B250000 and
above) pushbutton (in). Connect oscilloscope calibrator CHAN 1 to TI CH1 through 10x
attenuator and 50 Ω feedthrough termination.
(2) Press and release 20 MHz BW/TRIG VIEW pushbutton to full bandwidth and
press VERT MODE CH1 pushbutton (in).
(3) Set TI VOLTS/DIV CH1 switch to 5m (10m for Tektronix, Type 475A).
(4) Use technique of step 5 below for TI settings and oscilloscope calibrator output
settings listed in table 14.
(5) Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust
amplitude for 5 divisions of vertical deflection on TI. Square wave will have aberrations not
more than those listed in table 14; if not, perform adjustment as listed in table 14.
14