TB 9-6625-2240-35
Table 4. Channel 2
Vertical Deflection Accuracy
Test instrument
Oscilloscope calibrator
VOLTS/DIV
Oscilloscope calibrator
Err display limits
switch settings
Voltage output
%
3
2
m
12
mV
3
51
m
30
mV
3
10
m
60
mV
3
20
m
120
mV
3
50
m
.3
V
3
.1
V
.6
V
3
.2
V
1.2
V
3
.5
V
3.0
V
3
1
V
6.0
V
3
2
V
12.0
V
3
5
V
30.0
V
3
10
V1
60.0
V
1For
Tektronix, Type 475A.
particular TI you are calibrating. Perform appropriate adjustments listed in table 5 for 6
divisions of vertical display on TI.
Parameter
Tektronix
Adjustments
(mV gain)
type
Serial number
2
475
Below B250000
CH 1 R125
CH 2 R225
2
475
B250000 and up
CH 1 R195
CH 2 R295
5
475
All SN's
CH 1 R165
CH 2 R265
5
475A
---
CH 1 R195
CH 2 R295
10
475A
---
CH 1 R165
CH 2 R265
Main Sweep Timing
NOTE
For oscilloscopes with digital multimeters attached, refer to
a. Performance Check
(1) Press VERT MODE CH1 pushbutton and set VOLTS/DIV switch to .5 V.
(2) Perform steps (a) through (d) below:
(a) Press oscilloscope calibrator MARKER pushbutton to illuminate green LED
and ensure oscilloscope calibrator output is set to 1.000 ms.
(b) Connect oscilloscope calibrator CHAN 1 to TI CH1 through 50Ω
termination.