TB 9-6625-2251-35
evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer model
Common name
Minimum use specifications
(part number)
Voltage output:
(MIS38938) John Fluke, Model
CALIBRATOR
Range: 10 mV to 5 V
5820A (5820A-5C-GHz)
0.25%
Accuracy:
Time markers: 0.5 ns to .5 s
Leveled sine wave:
Range: 10 mV to 3 V p-p
Accuracy: 0.25%
CALIBRATOR
Voltage output:
John Fluke, Model 5720A (MIS-
Range: .000200 to 20 V dc
35947)
0.25%
Accuracy:
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 11-6625-3241-40.
d. Unless otherwise specified, all control and control settings refer to the TI.