TB 9-6625-2251-35
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUPUT(S) to minimum after each step within the performance
check where applicable.
a. Ensure the POWER-ON-OFF pushbutton is in the OFF position (out).
b. Connect TI to appropriate power source.
c. Press POWER-ON-OFF pushbutton to ON (in) and allow 20 minutes for TI to
warm up and stabilize.
d. Press MENU OFF/EXTENDED FUNCTIONS pushbutton once or twice.
NOTE
Two presses are necessary if any menu is presently displayed
and one press if no menu is displayed.
NOTE
"Select" means to press appropriate front panel pushbutton to
obtain a menu on crt. "Set" means to press appropriate menu
pushbutton located on lower portion of crt bezel by underlining
(turn on) or removing underline (turn off) desired menu
selection.
e. Set: CAL/DIAG.
f. Set: SELF CAL.
NOTE
RUNNING will be displayed in lower right corner of crt screen
for approximately 10 seconds. Upon successful completion of
self calibration, PASS will be displayed above SELF CAL
menu pushbutton.
g. Press MENU OFF/EXTENDED FUNCTIONS pushbuttons.
h. If TI displays EXTENDED DIAGNOSTICS menu, repeat d through g above. If TI
displays EXTENDED DIAGNOSTICS menu again, corrective action must be taken before
continuing procedure.
i. Select: STORAGE ACQUIRE pushbutton and verify NORMAL is on; if not, set:
NORMAL to on. Set: REPET ON:OFF pushbutton to ON.