TB 9-6625-2251-35
Table 14. MODE A Horizontal Linearity Accuracy
Test instrument
A and B
Display
MARKER
SEC/DIV
tolerance
output
setting
(major divisions)
50ms/div
50ms/div
0.1
.1s/div1
100 ms/div
0.1
0.1
.2s/div
200 ms/div
.5s/div
500 ms/div
0.1
1Select:
TRIGGER MODE and set: NORMAL to on.
g. Select: HORIZONTAL MODE B and set: TRIGGER MODE to AUTO.
h.
Use technique of
b through f above for settings listed in
table 15.
Table 15. MODE B
Horizontal Linearity Accuracy
Test instrument
A AND B
Display
MARKER
SEC/DIV
tolerance
output
setting
(major divisions)
500 ns/div
0.5 s/div
0.1
1s/div
1s/div
0.1
2s/div
2s/div
0.1
5s/div
5s/div
0.1
10s/div
10s/div
0.1
20s/div
20s/div
0.1
50s/div
50s/div
0.1
.1ms/div
100 s/div
0.1
.2ms/div
200 s/div
0.1
.5ms/div
500 s/div
0.1
1ms/div
1ms/div
0.1
2ms/div
2ms/div
0.1
5ms/div
5ms/div
0.1
10ms/div
10ms/div
0.1
20ms/div
20ms/div
0.1
50ms/div
50ms/div
0.1
.1s/div1
100 ms/div
0.1
.2s/div
200 ms/div
0.1
.5s/div
500 ms/div
0.1
1Select:
TRIGGER MODE and set: NORMAL to on
i. Set oscilloscope calibrator to
standby and
disconnect equipment setup.
13