TB 9-6625-2251-35
m Select: TRIGGER SOUCE and set: EXT then set: SOURCE 1⏐2 to 1. Connect
unconnected end of BNC input coupler to EXT TRIG 1.
in the first row of table 12. Rotate HORIZONTAL A AND B SEC/DIV knob for an
indication on crt as listed in the first row of table 12. Rotate TRIGGER LEVEL knob and
verify TI will trigger.
o. Repeat technique of n above substituting frequencies, amplitude, and TI setting in
Oscilloscope calibrator LEVEL SINE
Test instrument
SEC/DIV
Amplitude
settings
50
kHz
50
mV
5 s
80
kHz
50
mV
5 s
50 MHz
50 mV
5 ns
150 MHz
150 mV
5 ns
p. Move BNC dual input coupler connection from EXT TRIG 1 to EXT TRIG 2 and
perform (1) through (3) below.
(1) Select: TRIGGER SOURCE.
(2) Set: EXT
(3) Set: SOURCE 1⏐2 to 2.
q. Repeat technique of
n above using values in
table 13.
r. Set oscilloscope calibrator to
standby and
disconnect equipment setup.
Oscilloscope calibrator
Test
LEVEL SINE
instrument
SEC/DIV
Amplitude
settings
50
kHz
50
mV
5 s
80
kHz
50
mV
5 s
50 MHz
50 mV
5 ns
150 MHz
150 mV
5 ns
r. Select: SETUP AUTO.
11