TB 9-6625-2251-35
b. Rotate CH 1 VOLTS/DIV knob for 50 mV indications on crt display.
d. Select: SETUP AUTO.
NOTE
After auto setup sequence is completed, if necessary, adjust CH
1 VOLTS/DIV control for 50 mV indications on crt and
TRIGGER MODE to NORMAL.
e. Adjust oscilloscope calibrator knob located below the EDIT FIELD key on the
oscilloscope calibrator for 4 divisions peak-to-peak signal displayed on TI crt.
f. Rotate CH 1 VOLTS/DIV knob for 200 mV indications on crt. Rotate TRIGGER
LEVEL control and verify TI will trigger.
g. Repeat technique of b through f above substituting frequencies in the next 2 rows of
h. Rotate CH 1 VOLTS/DIV knob for 50 mV indications on crt display.
in the last row of table 11. Rotate TI TRIGGER LEVEL knob as necessary until TRIG'D
indicator illuminates.
j.
Select: SETUP AUTO.
NOTE
After auto setup sequence is completed, if necessary, rotate CH
1 VOLTS/DIV knob for 50 mV indications on CRT.
k. Select: STORAGE ACQUIRE and set: REPET to on. Adjust oscilloscope calibrator
knob locate below the EDIT FIELD key for 6 divisions peak-to-peak signal displayed on TI
crt.
l. Rotate CH 1 VOLTS/DIV knob for 200 mV indications on crt. Rotate TRIGGER
LEVEL knob and verify TI will trigger.
Oscilloscope calibrator
LEVEL SINE
LEVEL SINE
amplitude
frequency
200 mV
50 KHz
200 mV
80 KHz
200 mV
50 MHz
300 mV
150 MHz
10