TB 9-6625-2263-35
(3) Set calibrator for a 11 V, 20 kHz output. Adjust AC CONVERTER R425 10 V
20 kHz (fig. 1) for TI indication between 10.9988 and 11.0012 VAC (R).
(4) Set calibrator for a 11 V, 200 Hz output. Adjust AC CONVERTER R422 10 V
200 Hz (fig. 1) for TI indication between 10.9994 and 11.0006 VAC (R).
(5) Set calibrator for a 100 V, 200 Hz output. Adjust AC CONVERTER R433 100
V 200 Hz (fig. 1) for TI indication between 99.994 and 100.006 VAC (R).
(6) Set calibrator for a 100 V, 20 kHz output. Adjust AC CONVERTER R426 100
V 20 kHz (fig. 1) for TI indication between 109.988 and 110.012 VAC (R).
(7) Set calibrator for a 600 V, 200 Hz output. Adjust AC CONVERTER R438 600
V 200 Hz (fig. 1) for TI indication between 599.94 and 600.06 VAC (R).
a. Performance Check
NOTE
Do not perform power supply checks if all other parameters
are in tolerance.
(1) Connect multimeter INPUT HI to TP201 VCC and LO to TP202 VSS (fig. 3).
If multimeter does not indicate between +4.9 and +5.2 V dc, perform b(1) below.
(2) Connect multimeter INPUT HI to TP701 and LO to TP702 (fig. 3).
If
multimeter does not indicate between +4.9 and +5.2 V dc, perform b(2) below.
R204
R710
TP702
TP202
TP701
TP201
VSS
ANALOG CONTROLLER
VCC
AC
CONVERTER
Figure 3. Power supply -
adjustment locations.