TB 9-6625-2263-35
b. Adjustments
(1) Adjust R204 (fig. 3) for a multimeter indication between +4.9 and +5.2 V dc
(R).
(2) Adjust R710 (fig. 3) for a multimeter indication between +4.9 and +5.2 V dc
(R).
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
13