TB 9-6625-2266-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Digital Multimeter, Fluke, Models 8020A and 8020B. The manufacturers'
manuals were used as the prime data sources in compiling these instructions. The
equipment being calibrated will be referred to as the TI (test instrument) throughout this
bulletin.
a. Model Variations. Variations among models are described in text and tables.
b. Time and Technique. The time required for this calibration is approximately 1
hour using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
Performance specifications
parameters
Fluke, Model 8020A
Dc voltage
Range: 0 to 1000 V (in 5 ranges)
Accuracy: (0.25% of reading + 1 digit)
Ac voltage
Range: 0 to 750 V (in 5 ranges)
Accuracy: (% of reading + digits)
Range
Frequency
45 Hz to 1 kHz
1 to 2 kHz
2 to 5 kHz
200
mV
0.75 + 2
1.5 + 3
5.0 + 5
2
V
20
V
200
V
0.75 + 2
1.5 + 3
N/A
750
V
N/A
N/A
1.0 + 2