TB 9-6625-2266-24
(4) Repeat technique of (2) and (3) above using settings listed in table 3 below. If
TI does not indicate within limits specified in table 3, perform b below.
Calibrator
Test instrument
Indication limits1
Range
Output
setting
Min
Max
190
mV
200 mV
189.4 (189.7)
190.6
(190.3)
190
189.4 ( 189.7)
190.6 ( 190.3)
200 mV
mV
1.9 V
2V
1.894 (1.897)
1.906
(1.903)
19
V
20 V
18.94 (18.97)
19.06
(19.03)
190
V
200 V
189.4 (189.7)
190.6
(190.3)
1000
V
1000 V
997
(998)
1003
(1002)
Values in parenthesis apply to model 8020B.
1
b. Adjustments. Set TI range to 200 mV and calibrator for a 190 mV dc output.
Adjust DC CAL R6 (fig. 1) for a TI indication of 190.0 (R).
a. Performance Check
(1) Press DC/AC pushbutton to AC (in).
(2) Press TI 200 mV range pushbutton.
(3) Set calibrator for an output amplitude of 190 mV and an output frequency of
45 Hz. If TI does not indicate within limits specified in first row of table 4, perform b below.
6