TB 9-6625-2266-24
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287 and AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage:
Fluke, Model 5720A (5720A)
Range: 0.19 to 1000 V
(p/o MIS-35947); w amplifier,
Accuracy: (%)
Fluke 5725A/AR (5725A/AR)
8020B
8020A
0.19 to 190 V
0.079
0.039
1000 V
0.075
0.050
Ac voltage:
Range: 190 mV to 750 V
Accuracy: (%)
Voltage
40Hz and 1.0 kHz
190 mV through 190 V
0.211
2.0kHz
190 mV through 190 V
0.408
5.0 kHz
190 mV through 190 V
1.316
50 Hz and 1.0 kHz
750 V
0.317
Range: 190 to 19 M
Accuracy: (%)
8020B
8020A
190
0.115
0.090
1.9 through 190 k
0.063
0.038
1.9 M
0.063
0.513
19 M
0.513
0.513
Dc current:
Range: 1.9 mA to 1.9 A
Accuracy: 0.197%