TB 9-6625-2268-35
Test instrument
Calibrator output settings
Test instrument indications
range pushbutton
dBm
settings
(kHz)
Min
Max
200
mV
-55
1
-54.50
-55.50
200
mV
-20
1
-19.85
-20.15
2
-10
1
-9.85
-10.15
20
+15
1
+14.85
+15.15
200
+40
1
+39.85
+40.15
750V
AC
+45
1
+44.85
+45.15
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI mA and COMMON.
(2) Press and release AC/DC pushbutton to DC.
(3) Press mA function pushbutton.
(4) Press TI range pushbutton and set calibrator output to settings listed in table
6. TI
indications will be within limits
specified.
Test instrument
Calibrator
Test instrument
range pushbutton
output
indications
settings
settings
Min
Max
A
190 A
200
189.41
190.59
2
1.9 mA
1.8941
1.9059
20
19
mA
18.941
19.059
200
190 mA
189.41
190.59
2000
1.9 A
1894.1
1905.9
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI V/kΩ /S and COMMON.
(2) Press kΩ function pushbutton.
(3) Press TI range pushbutton and set calibrator output to settings listed in table
7. At
each setting, use calibrator output
adjustment controls to
set calibrator control
display Reading equal to TI indication. Calibrator control display Error indications
will be within limits specified in table 7.