TB 9-6625-2268-35
Test instrument
Test instrument
range
Indications
pushbutton
Calibrator output
settings
settings
Min
Max
Adjustments
200
mV
190
mV
189.92
190.08
b(1)
200
mV
-190
mV
-189.92
-190.08
2
1.9 V
1.8992
1.9008
b(2)
2
-1.9 V
-1.8992
-1.9008
20
19
V
18.992
19.008
200
190
V
189.92
190.08
b(3)
1000V
DC
1000
V
999.5
1000.5
b(4)
b. Adjustments
(1) Adjust R12 (fig. 1) for a TI indication of 190.00 (R).
(2) Adjust R11 (fig. 1) for a TI indication of 1.9000 (R).
(3) Adjust R5 (fig. 1) for a TI indication of 190.00 (R).
(4) Adjust R6 (fig. 1) for a TI indication of 1000.0 (R).
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI V/kΩ /S and COMMON.
(2) Press AC/DC pushbutton to AC.