TB 9-6625-2273-40
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Reference Calibration
Standards Set, NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and must
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI. Where the four-to-one ratio cannot be met, the actual
accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Precision termination, 50
BNC feedthrough,
0.1%, dc to 100 kHz,
Tektronix, Type 011-0049-01, and dc blocking capacitor, Tektronix, Type 015-0221-00.
Table 2. Minimum Specifications of Equipment Required
Common name
Minimum use specifications
Manufacturer and model
(part number)
FREQUENCY COUNTER
Frequency:
Fluke, Model PM6681/656
Range: 10 Hz to 500 MHz
(PM6681/656)
Accuracy: 0.25 PPM
MEASURING RECEIVER
Measuring receiver system N5530S
Range: 18.6 mV to 1.83 V
consisting of: Spectrum Analyzer,
Agilent Model E4440A (E4440A),
Accuracy: 0.375% 100 kHz to 100 MHz
Power meter, Agilent Model E4419B
Accuracy: 0.75% > 100 to 250 MHz
(E4419B), and Sensor module,
Accuracy: 1.0% > 250 to 500 MHz
Agilent Model N5532A opt. 504
(504)
Ac voltage:
Hewlett-Packard, Model 3458A
Range 18.5 mV to 1.96 V
(3458A)
Frequency: 20 Hz to 100 kHz
Accuracy: 0.375%
Range: 1.5 to 10 mV
Frequency: 50 kHz
Accuracy: 1.25%
Dc Voltage:
Range: 0 to 3 mV
Accuracy: 0.25%
SPECTRUM ANALYZER
Range: 20 Hz to 5 MHz
Hewlett-Packard, Model 3585B
NO. 1
(MIS-35951)
Level range: -55 dBm to + 20 dBm
SPECTRUM ANALYZER
Range: 10 MHz to 2.5 GHz
(AN/USM-677)
NO. 2
Level range: -40 dBm to + 20 dBm