TB 9-6625-2274-40
CALIBRATION PROCESS FOR PROGRAMMABLE
PULSE HEAD TEKTRONIX, TYPE 015-0611-00
23. Preliminary Instructions
a. The instructions outlined in paragraphs 23 and 24 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove TI bottom protective cover, as required, for adjustments.
c. Connect programmable calibration generator to a 115 V ac source.
warm-up and stabilization.
a. Performance Check
(1) Connect OUTPUT to digitizing oscilloscope using PULSE HEAD and X10
attenuator. (Use a different Xl0 attenuator if not in tolerance.) Connect TRIGGER
(2) Press pushbuttons as listed in (a) through (g) below:
AMPLITUDE MODE AMPL MODE FAST EDGE.
(a)
AMPLITUDE MODE EDGE POLARITY .
(b)
to highlight 100 kHz.
AMPLITUDE MODE FREQ
(c)
VARIABLE ON off.
(d)
TRIGGER ON on.
(e)