Time Markers - Continued
Fast Edge Pulse Risetime

Calibration Procedure For Programmable Calibration Generator Type Cg-5011 And Programmable Pulse Head Type 015-0611-00 Page Navigation
  18    19    20    21    22  23  24    25    26    27  
img
. . . . . .
TB 9-6625-2274-40
SECTION V
CALIBRATION PROCESS FOR PROGRAMMABLE
PULSE HEAD TEKTRONIX, TYPE 015-0611-00
23. Preliminary Instructions
a. The instructions outlined in paragraphs 23 and 24 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual.
d. Unless otherwise specified, all controls and control settings refer to the TI.
24. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Remove TI bottom protective cover, as required, for adjustments.
b. Connect TI to programmable calibration generator OUTPUT.
c. Connect programmable calibration generator to a 115 V ac source.
d. Energize programmable calibration generator and allow at least 30 minutes for
warm-up and stabilization.
25. Fast Edge Pulse Amplitude Mode
a. Performance Check
(1) Connect OUTPUT to digitizing oscilloscope using PULSE HEAD and X10
attenuator.  (Use a different Xl0 attenuator if not in tolerance.)  Connect TRIGGER
OUTPUT to digitizing oscilloscope TRIG.
(2) Press pushbuttons as listed in (a) through (g) below:
AMPLITUDE MODE AMPL MODE FAST EDGE.
(a)
AMPLITUDE MODE EDGE POLARITY  .
(b)
to highlight 100 kHz.
AMPLITUDE MODE FREQ
(c)
VARIABLE ON off.
(d)
TRIGGER ON on.
(e)
23